Edinburgh Instruments Ltd.
Are you Edinburgh Instruments Ltd. representative? Claim This Page Today!
No one from Edinburgh Instruments Ltd. has updated the information yet
Claim Edinburgh Instruments Ltd. Page to edit and add data


Frequently Asked Questions

The CO2 laser output is coupled into the FIR laser in the FIRL100 laser via two steering mirrors and a ZnSe focusing lens. Access to the CO2 radiation beam diagnostics for Infrared experiments is available via a precision, two position sliding mirror mechanism.

The resonator design of the FIRL100 laser is based on the proven PL5 laser with diffraction grating, two ZnSe Brewster windows and piezo ceramic mounted ZnSe output coupler.

The FIRL100 laser is a highly efficient optically pumped FIR system that combines both the CO2 pump laser and the FIR laser in a single compact unit. It is used for Infrared experiments.

The mode performance of the FIRL100 laser is assured by internal profiling of the tube and the use of high quality optics, with M2 <1.25.

The CO2 section of the FIRL100 laser provides 80 lines between 9.1μm and 10.9μm and features a flowing gas single discharge tube giving more than 50W on the strongest lines.