Frequently Asked Questions

Model SL 02/1 has one single mode, while Model SL 02/2 has two polarized modes.

The frequency stability over 1 minute / 1 hour / 24 hours relatively or absolutely after 40 minutes warm-up is approximately ± 1 / ± 2.5 / ± 5 MHz.

The laser safety class according to EN 60825-1 and ANSI Z136.1 (CDRH) is 3R IIIa.

Only a single 13.5 V-AC adapter is necessary for the power supply.

The warm-up time to achieve stable operation is less than or equal to 10 minutes.